23/09/2026
IOT-485-pH Digital pH Sensor | Calibration Guide
Accurate pH measurement is essential for water treatment, chemical processing, and industrial process control. In this video, we show the calibration procedure for the IOT-485-pH digital pH sensor.
The IOT-485-pH is a digital pH sensor with RS485 communication, designed for continuous online monitoring and IoT integration. It supports temperature compensation and menu-driven calibration for fast on-site maintenance. The calibration process includes preparation, sensor connection check, entering calibration mode, zero calibration with pH 7.00 standard solution, slope calibration with pH 4.00 or pH 10.00 standard solution, calibration verification, and returning to online measurement.
After calibration, the sensor provides stable pH readings and digital output for integration with PLC, DCS, SCADA, or IoT platforms. Typical applications include wastewater treatment, drinking water, surface water, industrial process water, aquaculture, and smart water monitoring systems.
More calibration videos and operation guides are available on our channel. This video is intended for water quality, maintenance, and process engineering teams.
https://www.youtube.com/watch?v=y2TM1RbGryg